[tomographie/includes/pelogo.htm] [tomographie/includes/bar.htm]
474
Timing Filter Amplifier
[tomographie/includes/horzline.htm]
474-fp.jpg (14276 bytes)
 
474-rp.jpg (9802 bytes)
  • Timing with germanium detectors
  • Energy spectroscopy at ultra-high count rates
  • Selectable filter for pulse shaping
  • Signal-to-noise ratio optimization
  • Continuously adjustable gain X2 to X250
  • Pole-zero cancellation
  • DC-coupling
The Model 474 Timing Filter Amplifier is especially designed to shape pulses and permit optimizing the signal-to-noise ratio for timing measurements. The Model 474 is particularly suited for use with an ORTEC Constant-Fraction Timing Discriminator in timing applications with germanium or silicon charged-particle detectors (Fig. 1). It derives its input signal directly from the preamplifier output. The timing spectrum in Fig. 2 illustrates the performance obtainable with the Model 474 shaping the germanium detector pulses before they are furnished to the discriminator. Tables 1 and 2 give typical performance data for various ORTEC germanium detectors.
The fast rise time, high output drive, and high gain capabilities of the Model 474 make it useful for other applications, such as timing with systems utilizing low-gain photomultiplier tubes. In addition, the pole-zero cancellation network, the dc-coupling, and the time-invariant baseline restorer permit energy spectroscopy with scintillation detectors and Si charged-partical detectors at ultra-high count rates.
A wide variety of input pulse shapes can be filtered as required for optimum signal processing. The Model 474 combines continuously adjustable gain (X2 to X250) with separately selectable Integrate (ti) and Differentiate (td) time constants for proper pulse shaping, making this unit and important asset for time measurement.

PERFORMANCE

Input Amplitude Range  0 to ±1 V signal, 0 to ±5 V dc offset; maximum input ±5 V total.
Output Amplitude Range  0 to ±5 V with a 50-W load.
Noise  For maximum gain, rms noise referred to the input is £10 µV with ti = td = 200 ns or £50 µV with filter out; measured using a Hewlett-Packard 3400A true rms meter.
Rise Time  £10 ns with filter out  or ~2.2 ti for other selections.
Nonlinearity  £±0.05% at midband frequency over ±5 V range.
Temperature Instability
DC Level  £±25µV/°C referred to the output.
Gain  £±0.06%/°C.
Specifications over 0 to 50°C range.

ELECTRICAL AND MECHANICAL

Power Required  +24 V, 65 mA; -24 V, 45 mA; +12 V, 160 mA; -12 V, 180 mA.
Dimensions  NIM-standard single-width module 3.43 x 22.13 cm (1.35 x 8.714 in.) per DOE/ER-0457T.
Weight
Net  1.0 kg (2.4 lb)
Shipping  2.5 kg (5.4 lb)

CONTROLS

COARSE GAIN  Front-panel 6-position switch for selection of X1, X2, X4, X6, X10, or X20.
FINE GAIN  Front-panel single-turn potentiometer, continuous from X2 to X12.5.
PZ ADJ  Front-panel screwdriver adjustment to compensate for the preamplifier decay time constant.
Time Constant  Two 6-position switched on front panel:
INTEGRATE  RC time constants: OUT (equivalent to 4 ns), 20, 50, 100, 200, and 500 ns.
DIFF  RC time constants: OUT (equivalent to 0.2 ms), 20, 50, 100, 200, and 500 ns.
NOTE:  With Differentiate and Integrate in the OUT position, the passband is 1 kHz to 35 MHz.
NON INV/INV  Selects inversion or non-inversion of the input signal.

INPUT

Positive or negative polarity selectable by front-panel switch; amplitude 0 to ±1 V; protected to ±6 V dc and to ±100 V at 10% duty factor integrated over 1 s; impedance 100 W, dc-coupled; front-panel BNC connector. Accepts a ±5 V dc offset, maximum input (signal plus offset) limited to ±5 V.

OUTPUTS

OUTPUT  Front-panel BNC connector.
Amplitude 0 to ±5 V; rise time £10 ns for filter out (2.2 ti for other filter selections). Impedance <1 W, dc-coupled.
Preamplifier Power  Rear-panel standard ORTEC power connector, Amphenol 17-10090.

Table 1.  Timing Resolution for Various Sizes of Germanium Detectors Using 22Na.

   Detector   Dynamic
Range
Timing Resolution (ns)
CF Mode SRT Mode
FWHM FW.1M FW.02M FWHM FW.1M FW.02M FW.01M
8.6% HPGe
52.6 cc
1.1:1
10:1
20:1
4.4
4.2
4.7
10.1
13.6
13.5
4.3
4.2
4.7
10.1
10.5
12.8
21.6
23.4
30.4
12.5% HPGe
62.3 cc
1.1:1
10:1
20:1
5.0
4.5
5.1
10.0
13.2
14.3
5.0
4.4
5.0
9.5
9.4
12.0
17.6
17.8
24.8
19.6% HPGe
103 cc
1.1:1
10:1
20:1
7.9
8.4
8.4
16.4
24.0
26.0
8.1
7.9
8.4
16.0
17.0
23.0
27.3
30.0
40.0
8.6% HPGe 1.1:1
5:1
10:1
6.4
7.6
7.6
12.6
18.0
22.7
29.1
59.4
63.2
6.5
7.3
7.6
14.1
18.0
21.7
29.6
45.6
50.1
  

Table 2. Timing Resolution for Large Germanium Detectors Using 583 CFDD/SCA, 474 TFA, and 60Co.

Detector Efficiency FWHM Energy Resolution (keV) Constant
Fraction
Delay
(ns)
Timing Resolution (ns)

E > 100 keV

E = 1332 ±50 keV
FWHM FW.1M FWHM FW.1M
N30526A 73% 2.03 34 5.4 19.4 3.7 8.8
P20171 81% 1.97 34 5.5 27.0 4.7 13.8
N20366A 88% 2.34 36 5.8 21.2 5.5 16.4
    
Fig. 1  Simplified Timing System. 474-f1.jpg (15555 bytes)
  
Fig. 2  Timing Spectrum for a Narrow Synamic Range Using a Germanium Detector. (Resolution values are given in Table 1.)

22Na
Start:  KL236 (1x1), RCA 8575 Photomultiplier Tube
Stop:  Ge Coax, 12.5%, 62.3 cc
1.1:1 Dynamic Range
473A Discriminator

474-f2.jpg (25894 bytes)